search for


Concurrent Validity and Clinical Usefulness of Universal Plastic Goniometer for Hip Internal and External Rotation Range Measurement
J Korean Soc Phys Med 2018;13(1):99-105
Published online February 28, 2018;
© 2018 Korean Society of Physical Medicine.

Yong-Wook Kim, PhD, PT

Dept. of Physical Therapy, College of Medical Sciences, Jeonju University
Received January 10, 2018; Revised January 14, 2018; Accepted February 7, 2018.
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License ( which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
PURPOSE: The aim of this study was to evaluate the concurrent validity and clinical usefulness of the universal plastic goniometer to measure the range of motion of the internal and external rotation of the hip joint using the three dimensional motion analysis which can analyze the joints and segment movements in the most objective and quantitative method.
METHODS: Clinical and kinematic data were collected from thirty individuals using a universal plastic goniometer and a ten camera motion analysis system. Passive hip rotation range was obtained three trials for left and right hip joints using two measure methods simultaneously.
RESULTS: There were significant differences between all matching measures of the two measures of internal and external rotation of the hip joint (p<.05). The relationship between the two tests for all measurements of the internal and external rotation of the hip was statistically significant with correlation coefficient form r=.87 to .96. (p<.01).
CONCLUSION: Clinical measurement of the internal and external rotation of the hip using a universal plastic goniometer is effective to assess the hip condition. However, application of universal plastic goniometer requires careful attention in more accurate evaluation and research verification of the internal and external rotation of hip joint.
Keywords : 3D motion analysis, Goniometry, Hip joint, Range of motion

February 2018, 13 (1)
Full Text(PDF) Free

Social Network Service

Cited By Articles